42
Transmit or Receive operation.
Command: /Dxx[CR] 'Set Destination I
2
C Slave Address
Response 1: * 'iPort/AFM Ready
Response 2: /I89[CR] 'Invalid Command Argument
Default Setting: 00
Echo/Prompt Control
This command enables or disables data entry echo and prompts used as feedback to
manual operations from a computer terminal.
Command: /E[0|1][CR] 'Echo/Prompt Control [0 = Off, 1 = On]
Response: * 'iPort/AFM Ready
Default Setting: Off
RS-232 Flow Control
Select the serial communication handshaking protocol to be use in communicating
with the Host computer.
iPort/AFM implements either XON/XOFF (by default) or RTS/CTS flow control
protocols. Flow control is used by the iPort/AFM to limit character flow to and
from the Host computer to avoid overflowing internal communication buffers and
lost data.
The XON/XOFF protocol inserts characters directly into the ASCII data stream.
XON (Hexadecimal 0x11) is used to enable the flow of data. XOFF (Hexadecimal
0x13) is used to stop the flow of data.
The RTS/CTS protocol uses two additional wires in the cable connecting
communicating devices. The RTS wire is an output signal. It indicates that the
device generating the signal has buffer space available, and can receive data. The
CTS wire is an input signal. It indicates that the other device has buffer space
available, and can receive.
In general, XON/XOFF requires a minimal three-wire connection, Ground,
Transmit Data, and Receive Data. This protocol does insert control characters into
the stream of data, and may not be appropriate for all Host systems. If supported,
these control characters are normally automatically stripped out of the data stream